MPN: AD 200
PHOTOMETRIC ACCURACY (ELECTRONICS/SOFTWARE) 1% and +0.005 OD to 2 OD
POWER USAGE
Fuses, a (MEASUREMENT) 2.5
POWER NEEDED, VAC, HZ (TEST CAPABILITY) 100-240, 50-60
Usage, w (MEASUREMENT) 250 VA (max)
DIMENSIONS (HXWXD) CM, (IN) (DISPLAY) 26 x 48 x 44 (11 x 19 x 18)
WEIGHT, kg (lb) (DISPLAY) 17 (37)
PHOTOMETRIC TECHNIQUE (MEASUREMENT) Endpoint kinetic - for details please see comments multiwavelength, linear and area-position scanning, spectral scanning
Precision (WAVELENGTH, nm) 1% and +0.005 OD to 2 OD
RESOLUTION, OD (ELECTRONICS/SOFTWARE) 0.001
TEMPERATURE
Operating range, c (TEMPERATURE) 15-40, 15-75% humidity
Incubator operation range, c (TEMPERATURE) -25 to +50
PHOTODETECTOR Silicon photodiode
FILTER USED (ASPIRATOR) Grating monochromator
RECEPTACLE (POWER USAGE) 6- to 384-well
Measurement
Range, abs (MEASUREMENT) 00.0-4
WAVELENGTH RANGE, nm (UV RADIATION INTENSITY) 190-1,000
ISOTHERMAL READING CHAMBER (ELECTRONICS/SOFTWARE) Yes
C (TEMPERATURE) Temperature control up to 45
ELECTRONICS/SOFTWARE (NO SOFTWARE INCLUDED)
PRINTER INTERFACE Centronics parallel
Programmable tests (ELECTRONICS/SOFTWARE; NO SOFTWARE INCLUDED) Unlimited (PC)
Reading speed, sec/plate (ELECTRONICS/SOFTWARE; NO SOFTWARE INCLUDED) 11 sec/96-well, 22 sec/384-well
PRINTER (EXTERNAL BEAM) PC printer
INTERFACE PROTOCOL (REPORT DISPLAY) RS232
Methods of calculation (ELECTRONICS/SOFTWARE) Standard curves, formula calibration, groups (cutoff), validation, replicate control, variables
Equivalent microplates (ELECTRONICS/SOFTWARE) All plates from 6- to 384-wells with transparent bottom
ILLUMINATION SOURCE Xenon flashlamp
OPTICAL SYSTEM (SPECIAL/TIME, min) Absorbance only
MARKETING REGION (Interference compensation) Worldwide
Return Policy
30 Day Return Policy. Free Return. One month free returns for items sold within the United States. No returns for international shipments.