This item has been biomed tested and verified to be in good physical and working condition. 30 Day Warranty.
Arthrex O.P.E.S. Electrosurgical Generator
REF: AR-9600
(Similar to Bovie IDS-300)
Includes: Bovie Monopolar Foot Switch BV-1253B
Bipolar footswitch NOT INCLUDED.
POST: Pass
Buttons: Pass
Alarm Tone: Pass
Activation Tone: **Fail**- Likely a faulty volume potentiometer. Tone temporarily functions when potentiometer is pressed inward with finger.
Display and indicator lights: Pass
Return Electrode Alarm: Functions
Footswitch ports for Monopolar and Bipolar: Function
Electrical Safety Test (NFPA992012): Pass
Power test and HF leakage test on Rigel Analyser:
Cut 1 @ 300 ohms-
Power setting 300: 271 W, 950 mA, Pass
Power setting 150: 138 W, 677 mA, Pass
Power setting 50: 46 W. 393 mA, Pass
HF leakage active electrode: 66 mA, Pass
HF leakage neutral electrode: 95mA, Pass
Cut 2 @ 300 ohms-
Power setting 300: 271 W, 950 mA, Pass
Power setting 150: 138 W, 678 mA, Pass
Power setting 50: 46 W, 392 mA, Pass
HF leakage active electrode: 66 mA, Pass
HF leakage neutral electrode: 94 mA, Pass
Blend @ 300 ohms-
Power setting 200: 186 W, 787 mA, Pass
Power setting 100: 95 W, 561 mA, Pass
Power setting 50: 46 W, 391 mA, Pass
HF leakage active electrode: 69 mA, Pass
HF leakage neutral electrode: 94 mA, Pass
Pinpoint Coag @ 500 ohms-
Power setting 120: 108 W, 465 mA, Pass
Power setting 60: 58 W, 342 mA, Pass
HF leakage active electrode: 665 mA, Pass
HF leakage neutral electrode: 93 mA, Pass
Spray Coag @ 500 ohms-
Power setting 80: 76 W, 391 mA, Pass
Power setting 40: 39W, 277 mA, Pass
HF leakage active electrode: 108 mA, Pass
HF leakage neutral electrode: 101 mA, Pass
Bipolar @ 150 ohms-
Power setting 80: 71 W, 688 mA, Pass
Power setting 40: 38 W, 502 mA, Pass
HF leakage 1st electrode: 22 mA, Pass
HF leakage 2nd electrode: 43 mA, Pass
Return Policy
30 Day Return Policy. Customer must pay the return shipping costs when returning a product. Items are sold as-is with no returns or refunds available unless explicitly stated.